1906

1906 - 5.5 Digit LED Bench Multimeter

  • 5½ digits (±210,000 counts), auto/manual range
  • 1µV, 1nA, 1mOhm resolution, 0.012% accuracy
  • Wide range of computing functions included
  • Automatic data logging functions included
  • Integral True RMS AC converter
  • Isolated addressable RS232 interface standard
  • Model with GPIB (IEEE-488) interface available

ACCURACY
Accuracies apply for 1 year 18oC to 26oC. Temp. coefficient outside these limits is <0×1 x quoted range
accuracy per oC. Warm-up time to rated accuracy is 1 hour.

DC VOLTS

Range

Accuracy(±)

Input Impedance

Max Input

200mV

0×017% + 3d

10M or >1G (Hi Z mode)

1kV DC or AC Peak any range   

2V

0×012% + 3d

 

20V

0×019% + 3d

10M

200V

0×019% + 3d

 

1000V

0×019% + 3d

 

RESISTANCE (OHMS)

Range

Accuracy(±)

Max. Measuring Current

Max. Input

200

0×025% + 4d

1×5mA

300V DC or RMS continuous any Ohms input any range

2k

0×019% + 3d

1×5mA

20k

0×019% + 3d

0×4mA

200k

0×019% + 3d

40mA

2M

0×022% + 3d

4mA

20M

0×07% + 3d

400nA

Max open circuit voltage 3×5V.  Diode Test  on 2k range.
Audible continuity check available on all ranges.

AC VOLTS (True RMS)

Range

Accuracy(±)

 

45Hz - 5kHz

5kHz - 10kHz

10kHz-20kHz

20kHz-50kHz

200mV

0×2% + 150d

0×2% + 250d

3% + 500d

-

2V

0×2% + 150d

0×2% + 250d

0×5% + 300d

3% + 1500d

20V

0×2% + 150d

0×2% + 250d

0×5% + 300d

2% + 1000d

200V

0×2% + 150d

0×2% + 250d

0×5% + 500d

-

750V

0×2% + 150d

-

-

-

Additional error at crest factor=3 typically 0×7%, Input impedance 1M / <100 pF any range.
Max input 750V rms, 1kV peak, any range.

DC AMPS

Range

 

Accuracy(±)

Voltage Burden

Max. Input

200mA

 

0×08% + 12d

300 mV max

1A, 300V fuse protected

2mA

 

0×08% + 12d

 

20mA

 

0×08% + 12d

 

200mA

 

0×08% + 12d

 

10A

 <2000mA

0×08% + 12d

650mV max

10A, 300V fuse protected

 

>2000mA

0×12% + 12d

 

AC AMPS (True RMS)

Range

Accuracy(±) 45 Hz - 1 kHz

Voltage Burden

Max Input

200mA

 

0×37% + 100d

300mV max

1A, 300V fuse protected

2mA

 

0×37% + 100d

 

20mA

 

0×37% + 100d

 

200mA

 

0×37% + 100d

 

10A

<2000mA

0×37% + 100d

650mV max

10A, 300V fuse protected

 

>2000mA

0×4% + 100d

 

Additional error at crest factor=3 typically 0×7%.

AttachmentSize